Walter F. Stafford BOSTON BIOMEDICAL RESEARCH INSTITUTE ANALTYICAL ULTRACENTRIFUGATION RESEARCH LABORATORY 64 GROVE STREET, WATERTOWN, MA 02472 Voice 617 - 658 - 7808 E-mail: stafford@bbri.org Revised November 12, 2000 3. Time interval and Maximum number of scans to analyze. A rule of thumb for choosing the maximum number of scans to analyze is the following: The time difference, delta(t), between the first and the last scan used should not exceed the following value: delta(t)max < [900*t]/[(M^0.5)*(SPEED)] where t is the average of the times for first and last scan. M is the molecular weight expressed in Kg/mole and SPEED is the angular velocity of the rotor expressed in units of (RPM/1000). This can also be expressed more handily as delta(w^2t)max < [90*(w^2t)]/[(M^0.5)*(SPEED)] where now w^2t is the value of "omega squared t" from the second line of the XL-A/XL-I output files. #This rule of thumb is based on the idea that the boundary, assumed to be gaussian and located in the middle of the cell, should not be allowed to move more than one standard deviation during the time interval over the subtraction is to carried out. Empirically, this rule avoids excessive artificial broadening. This is not a hard-and-fast rule. You should verify that it works in your case by trying smaller and larger numbers of files in your data set. In some cases you might be able to tolerate fatter peaks in order to get better signal-to-noise by including more scans. The peaks will be artificially broadened but the peak positions would be unaffected. For purposes of curve fitting to get the apparent diffusion coefficient, you might try a more stringent criterion, for example, delta(w^2t)max < [40*(w^2t)]/[(M^0.5)*(SPEED)]